Stability Chambers & Environmental Test Equipment
Reach-in and walk-in stability chambers, incubators, freezers and ovens for pharmaceutical stability studies.
Stability chambers: quick answers
What temperature and humidity range do Aureole stability chambers cover?
The Humidity Chamber and Stability Chamber are specified for 20°C to 60°C and 40% to 95% RH, with accuracy listed as ± 0.3°C / ± 2.0% RH. The Ultra Low Deep Freezer is specified for up to -80°C. Humidity Chamber specifications
Which standards are the photostability chambers designed for?
Aureole Photostability Chambers are described as designed in accordance with ICH Q1B guidelines, with light-intensity tracking alongside temperature control. Photostability Chamber
What is the difference between a stand-alone chamber and a walk-in chamber?
Stand-alone chambers are reach-in units. Walk-in chambers (humidity, cold, BOD incubator and deep freezer variants) are room-sized units for large-scale storage and testing. Walk In Humidity Chamber
Can chambers be monitored with software?
Yes. Aureole monitoring software supports 21 CFR Part 11 requirements through access controls, electronic signatures and audit trails, and also offers SCADA integration and SMS/email alerts. 21 CFR Part 11 support is listed as an add-on option on chamber product pages. Compliance of a given installation also depends on how the customer validates and operates it. Monitoring software